Medical & Diagnostic Equipment

Medical Monitor Housing: Conductive Seam Improvement

The housing protected a portable ECG monitoring device and formed part of its electromagnetic shielding arrangement.

Anonymous Customer Case Study

📅 July 7, 2026 ⏱ 2 min read 🏭 DRAmetal Engineering
Medical Monitor Housing: Conductive Seam Improvement — component reference
Component reference image
≤0.1 Ω Seam resistance
250 × 180 × 80 mm Housing envelope
1.0 mm Sheet thickness

Project Overview

Application The housing protected a portable ECG monitoring device and formed part of its electromagnetic shielding arrangement.
Materials 1.0 mm SECC steel with a conductive surface treatment.
Process Route Laser cutting → bending → welding → grinding → conductive finishing → screen printing → conductive-gasket assembly.
Customer Challenge Radiated-emissions testing identified leakage at enclosure seams where electrical continuity was inadequate.

The housing protected a portable ECG monitoring device and formed part of its electromagnetic shielding arrangement.

The customer’s challenge

Radiated-emissions testing identified leakage at enclosure seams where electrical continuity was inadequate.

Material and fabrication route

1.0 mm SECC steel with a conductive surface treatment.

Process route: Laser cutting → bending → welding → grinding → conductive finishing → screen printing → conductive-gasket assembly.

Manufacturing changes

The housing overlap was revised, gasket channels were added, and fingerstock contacts improved conductive continuity along the seams.

Engineering analysis

The original issue was conductive continuity at the enclosure seams. Revised overlaps, gasket channels and fingerstock contacts addressed that specific leakage path. The seam-resistance result and the configured device’s reported emissions acceptance answer different questions. For another monitor, the board, connector and enclosure interfaces need to be considered together with the customer’s test setup.

Inspection and project results

The device passed the customer’s reported EMC emissions test after the seam changes. The measured seam resistance and housing envelope are recorded below.

Project measurement or production detailRecorded value
Seam resistance≤0.1 Ω
Housing envelope250 × 180 × 80 mm
Sheet thickness1.0 mm

These figures describe the project configuration and production scope above. Dimensional, material and production details are identified separately from performance tests. For a new order, the inspection method and acceptance criteria should be agreed against the released drawings.

Planning a similar OEM component

A useful RFQ for this application should make the following interfaces and acceptance requirements clear:

  • Board and connector arrangement.
  • Conductive seam and gasket details.
  • Finish masking at contact surfaces.
  • Configured-device emissions acceptance.

Include the current drawing revision, mating-component information, quantities and required delivery sequence. Identifying unresolved requirements at quotation stage helps distinguish fabrication work from the equipment-level checks that the project needs.

For related manufacturing support, review our Medical Device Sheet Metal Housings and sheet metal quality capabilities. Use the existing quotation form below to share the drawing package for review.

More Manufacturing Projects

Explore More Manufacturing Work