A medical equipment customer needed a stainless tray assembly that could accommodate different surgical instrument sets while addressing edge damage, difficult corner access and residual watermarks after steam processing. The original fixed-divider arrangement restricted changes to the kit layout. Burrs around perforations also created an unwanted contact condition for the instruments.
DRAmetal’s project combined changes to the tray body, removable partitions and finished surfaces. The assembly used 1.0 mm 304 stainless steel for the tray, 0.8 mm 304 stainless steel for the dividers, and silicone locating pads. The result was an adjustable internal arrangement with controlled perforations, rounded welded corners and drainage feet.
The project recorded hole deviations of 0.02–0.04 mm, flatness of 0.15–0.25 mm/m, and no snagging in the glove-wipe check. Three validation samples completed 500 steam cycles at 121°C, with dimensional change ≤0.1% and no visible surface corrosion. The tray project also passed acceptance by the hospital’s central sterile supply department.
The customer’s problem connected handling, layout and processing
The perforations and internal partitions were not independent details. Instruments contacted the tray and locating features, while the tray needed to accommodate more than one kit arrangement. A fixed partition could be suitable for one set yet obstruct another, forcing the customer to reconsider the whole internal layout.
The supplied project information identified four problems:
- Burrs at punched openings could scratch instruments.
- Fixed dividers could not accommodate different instrument sets.
- Tray corners were difficult to access during cleaning.
- Watermarks remained after steam processing.
Addressing only the tray’s external dimensions would have left those issues unresolved. The manufacturing changes therefore covered the hole edges, divider attachment, corner construction and underside support arrangement as one component package.
Materials and recorded manufacturing route
The tray body and dividers used the same stainless grade but different thicknesses. Keeping these two drawing requirements distinct preserved the intended construction of the body and the removable internal parts. Silicone pads provided the specified locating interfaces within the assembly.
Recorded route: Laser cutting → bending → continuous welding → electropolishing → passivation → divider punching → silicone-pad assembly → cleaning → sterilization validation.
The project changes included deburring after punching and electropolishing of the specified stainless surfaces. That edge treatment was important because a dimensionally acceptable opening could still have an unacceptable contact edge. The acceptance plan therefore addressed both the dimensional characteristics and the finished condition.
For repeat procurement, the useful controlled package is the body drawing, divider drawing, pad reference and assembly configuration together. A substitute divider or pad can affect the tray even when the outside dimensions are unchanged. The instrument tray RFQ guide explains how to identify those parts before requesting a quotation.
Removable dividers replaced the fixed arrangement
The revised assembly used removable clip-connected dividers with adjustable spacing. This allowed the internal arrangement to be changed for different instrument sets without replacing the entire tray body.
The manufacturing change also made the attachment and removal interfaces part of the component review. The divider had to fit its intended locations while remaining removable for the customer’s required access. The design benefit came from the relationship between the divider, receiving features and locating pads, rather than from a change to the tray envelope alone.
For a comparable OEM program, permitted positions and orientations should be included in the assembly information. This project’s adjustable spacing demonstrates its chosen configuration; another instrument kit needs its own arrangement and acceptance review. The divider retention guide sets out the relevant drawing inputs.
Rounded welded corners and drainage feet
The corner construction changed to R ≥8 mm with continuous welding. This addressed the difficult access associated with the previous corner geometry. Drainage feet were added as part of the response to residual water after steam processing.
These changes treated the tray’s inside transition and supported position together. A rounded internal corner addresses one part of the geometry, while feet change how the underside rests against a supporting surface. Both belong in the released assembly definition when the tray is ordered again.
The project reported that the revised corners resolved the identified access problem. The R ≥8 mm dimension is the value used for this design, not a general rule for all medical trays. A new configuration should coordinate corner shape, seams, perforations and supports with the intended processing arrangement. See the drainage and corner review for procurement questions.
Inspection results and their coverage
The following results keep dimensional inspection, finished-edge checks and validation samples separate. Scroll horizontally on a small screen to view all columns.
| Characteristic | Project requirement | Recorded result | Recorded coverage |
|---|---|---|---|
| Hole tolerance | ±0.05 mm | Deviation 0.02–0.04 mm | First-article full inspection; five pieces per batch |
| Tray flatness | ≤0.3 mm/m | 0.15–0.25 mm/m | First-article full inspection; five pieces per batch |
| Finished perforation edges | Burr height ≤0.02 mm; no sharp or tactile edges | 100% deburring; glove wipe without snagging | 100% tactile inspection; sampling at 10× magnification |
| Steam-cycle durability | 121°C for 500 cycles without deformation or corrosion | Dimensional change ≤0.1%; no visible surface corrosion | Three validation samples |
The edge record describes the completed deburring and inspection observations. The 0.02 mm burr-height value is a drawing requirement, not a reported measured burr-height result. Keeping that distinction avoids treating a successful glove-wipe check as a numerical measurement.
Likewise, the three-sample steam evaluation demonstrates the recorded dimensional and surface durability of this configuration under the stated exposure. It does not establish sterility for every instrument load or provide a general service-life rating for every tray. The applicable processing procedure and its validation remain specific to the complete device and use configuration.
Outcome and relevance to a new OEM order
The customer received a tray configuration with adjustable divider spacing, finished perforation edges, accessible rounded corners and drainage feet. The recorded inspection and durability results supported this project, and the assembly passed the hospital central sterile supply department’s acceptance.
For a similar procurement, start with the intended kit layout and the released component drawings. Identify the tray, lid where required, divider positions, pads, finished-edge criteria and supported orientation. Then agree dimensional sampling and any separate validation activity before production.
Custom medical instrument trays can be quoted around that defined package. The first-article inspection guide helps translate the drawing requirements into results that purchasing and engineering can review without conflating a sample test with inspection of every production part.